Fluke 772 Milliamp Process Clamp Meter

Fluke771.772.773 Milliamp Process Clamp Meter

Fluke771.772.773 Milliamp Process Clamp Meter

Fluke 772 Milliamp Process Clamp Meter

Introduction:

This manual tells you about verification and adjustment procedures for the 772/773 Milliamp Process Clamp Meter (referred to in this manual as the Meter). The Meter features closed-case calibration to use with reference sources. It measures the reference signals, calculates the correction factors, and keeps them in memory. Calibration adjustment is required after a repair, or if the Meter fails a performance test.

Features:

Technical specifications
  Range Resolution Accuracy
AC current Via Jaw 0-20.99 mA; 21-100mA 0.01Ma; 0.1 mA 0.2 % ± 5 counts

1 % ± 5 counts

In Circuit 0-24 mA 0.01 mA 0.2 % + 2 counts
Current Source 0-24 mA 0.01 mA 0.2 % + 2 counts
mA Drive 24 mA into 1000 Ω
AC voltage 600 V/1000V 0.1 V/1V 1.5 % ± 5 digits (20-500 Hz)
DC voltage 600 V/1000V 0.1 V/1V 1 % ± 5 digits
Frequency Via Jaw 5.0-500.0Hz 0.1Hz 0.5 % ± 5 digits
Trigger Level 5-10 Hz, ≥10A

10-100 Hz, ≥5A

100-500 Hz, ≥10A

Frequency Via Flexible Current Probe 5.0-500.0Hz 0.1Hz 0.5 % ± 5 digits
Trigger Level 5-20 Hz, ≥25A

20-100 Hz, ≥20A

100-500 Hz, ≥25A

Resistance 600/6000Ω/60kΩ 0.1/1Ω/10Ω 1 % ± 5 digits
General Specifications
Size (LxWxH) 277 mm × 88 mm × 43mm
Jaw opening 34 mm
Weight 350 g
Flexible Current Probe Diameter 7.5mm
Flexible Current Probe Cable Length 1.8m
Safety rating CAT III 600 V
Warranty 2 years
Mechanical Specifications  
Batteries 2 AA, NEDA 15A, IEC LR6  
Operating temperature -10 °C to +50 °C  
Storage temperature -40 °C to +60 °C  
Operating humidity –
non-condensing (< 10°C)
≤90% RH (at 10 °C to 30 °C)  
≤75% RH (at 30 °C to 40 °C)  
≤45% RH (at 40 °C to 50 °C)  
Operating altitude 2000 m  
Storage altitude 12,000 m  

Include: