Fluke 771 Milliamp Process Clamp Meter

$689.00

Fluke 771 Milliamp Process Clamp Meter

Fluke 771 Milliamp Process Clamp Meter

Introduction:

This manual provides the complete verification and adjustment procedure for the 771 Milliamp Process Clamp Meter (referred to in this manual as the Meter). The Meter allows closed-case calibration using reference sources. It measures the reference signals, calculates the correction factors, and stores them in memory. The instrument should be calibrated after repair, or if it fails a performance test.

Specifications:

Current Ranges ±20.99 mA ±21.0 mA – ±99.9 mA
Resolution 0.01 mA 0.1 mA
Accuracy
20.99 mA range 0.2 % reading ±5 digits
99.9 mA range 1 % reading ±5 digits
Maximum Reading ±99.9 mA
Influence of Earth’s Field < 0.20 mA
Battery 2 AA 1.5 V Alkaline, IEC LR6
Working hours 45 hours
Size (H X W X L) 59 mm x 38 mm x 212 mm  (2.32 in x 1.49 in x 8.34 in) (with clamp nested)
Weight 260 g (9.17 oz) (Including battery)
Operating Temperature -10 to 50 ℃(14 to 122°F)
Storage Temperature -25 to 70  ℃ (-13 to 158°F)
Operating Humidity < 90 % @ <30 ℃ (86°F)
<75 % @ 30 to 50 ℃(86 to 122°F)
Operating Altitude 0 to 2000 m (1.24 miles)
Storage Altitude None
IP Rating IP 40
Vibration Requirements Random 2 g, 5 to 500 Hz
EMI, RFI, EMC Meets all applicable requirements in EN 61326-1
Temperature Coefficients 0.1x(specified accuracy)/°C

(< 18 °C or > 28 °C)

Measurement Category IEC 61010-1

61010-2-032

CAT II 300 V

Weight 0.26 kg
Dimensions 212 × 59 × 38 mm
Brand

Suppliers